Testing method to validate the cleaning process in terms of organic content on medical devices has been established following the proposed standard XP S 94-091.

Type of pollutionResearched elementsAcceptanceAnalysis method and references
Total hydrocarbons (THC)Hydrocarbon≤ 0,5 mg/implantGas chromatography (NF ISO 9377-2)
Total Organic Carbon (TOC)TOC≤ 0,5 mg/implantTOCmeter (NF EN 1484)

These tests are conducted to determine the absence of manufacturing oil, polishing grease, lubricant, etc. on orthopedic implants.